VACUUBRAND PC 600 LAN NT User Manual

Page 27

Advertising
background image

page 27 of 102

Controller

VNC 2

Measurement uncertainty (absolute)

after careful adjustment and at constant

temperature

<

± 0.75 Torr (1 mbar) ± 1 digit

Temperature coefficient

<

± 0.05 Torr/K (0.07 mbar/K)

Ambient temperature range (operation)

50°F to 104°F (10°C to +40°C)

Ambient temperature range (storage)

14°F to 158°F (-10°C to +70°C)

Permissible relative atmospheric mois-

ture during operation (no condensation)

30% to 85%

Maximum permitted range of supply

voltage

100 V~ (-10%) to 230 V~ (+10%)

50/60 Hz

Fuse (IEC connection)

slow-blow fuse 5x20 mm, T8A / 250 V

Power draw (no-load operation)

3 VA

Power draw with VACUUBRAND VV-B

6C in-line valve and coolant valve

max. 10 VA (without switch output)

Interface

RS-232 C

Supply line VACUU•BUS

voltage supply 110-230 V 50/60 Hz

voltage supply 100-110 V 50/60 Hz

designed for the simultaneous opera-

tion (parallel connection) with the fol-

lowing original accessories:

24 V, max. 400 mA

24 V, max. 340 mA

in-line valve VV-B 6C 24 V DC

coolant valve VKW-B 24 V DC

venting valve VB M-B 24 V DC

Breaking capacity of IEC socket

ambient temperature 86°F (30°C)

ambient temperature 104°F (40°C)

7.2(4)*** A 250 V~ / 7.2(7.2)*** A 125 V~

6(4)*** A 250 V~ / 6(6)*** A 125 V~

Degree of protection IEC 529

IP 20

Vacuum connection

connection for PTFE tube

O.D. 3/8”, I.D. 5/16” (10 / 8 mm)

Weight approx.

1.8 lbs. (0.8 kg)

Dimensions L x W x H approx.

(without connection for PTFE tube)

6.4” x 3.5” x 2.7”

(163 mm x 90 mm x 68 mm)

Mounting of stand rod

via thread M8

Power supply

integrated IEC combination

* The actual vacuum control range in your application might be reduced due to ultimate vacuum

of the pump, volume of gas present, etc.

** if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)

*** ohmic (inductive) load