VACUUBRAND PC 600 LAN NT User Manual

Page 53

Advertising
background image

page 53 of 102

not been reached (pressure increase control).

+

Upper pressure value (p On, factory-set: 150 mbar): Condition for

restart of the pump when the pressure is increasing. If the pressure

exceeds this pressure, the pump starts again.

+

Lower pressure value (p Set, factory-set: 25 mbar): Condition for

switching off the pump. If the pressure drops below ”p Set”, a time-

meter (see ”Switch-off time”) starts to run. If the vacuum is below ”p
set

” for longer than the ”Switch-off time”, the process control stops.

The vacuum pump is switched off.

+

Time of process (tProcess, factory-set: 99 h, only via serial inter-

face): Check if the switch-off time is reached within the preset time

of process. If the switch-off time is not reached, an error message

”CheckSys” appears. In that case, check the process parameters

and/or the vacuum system, see section “Troubleshooting”, pg. 77.

+

Preselection of

Auto-Start (Auto-St., factory-set ”Off”): Process

starts immediately after switching on the controller.

Attention: If the controller is set to Autostart (Auto-
St.: On

), the process will start immediately without

pressing any further key after a power failure, or after

switching on the controller if process control has not

been terminated by pressing ”Esc”. It is your respon-

sibility to ensure that automatic start-up of the system

will not lead to any dangerous condition. Provide ap-

propriate safety measures. Check prior to starting the

process whether the option ”Autostart” (device menu)

is enabled.

Suggested values for process parameters

Pumping unit

Switch-off time

(min)

Lower pressure

value (Torr/mbar)

Upper pressure

value (Torr/mbar)

PC 500 LAN NT

5-10

11-15 / 15-20

75-150 / 100-200

PC 600 LAN NT

5-10

4-8 / 5-10

75-150 / 100-200