VACUUBRAND PC 600 LAN NT User Manual

Page 91

Advertising
background image

page 91 of 102

Bring the diaphragms (F) into a position, in

which they are in contact with the housing

(C) and centered with respect to the bore.

Put on head cover (J) with valve heads (O)

and connections attached.

+

Pay attention to the correct orientation of

the head covers:

Housing with head alignment pin: The head

alignment pin (A) at the pump housing (C)

has to fit into the recess at the head cover

(J).

Housing with mark (A): Align the recess at

the head cover with the mark at the pump

housing.

Loosely screw in the Allen head screws (H)

at the head covers diagonally at first slightly

with a 5 mm wide Allen key, then tighten.

+

Recommended torque: 8.9 ft

.

lb

f

(12 Nm).

H

Affix the connection tube (V) to the other side

of the pump, as well as the hose connection to

the inlet or outlet of the vacuum system at the

valve head (O).

Slip connecting tube (V) onto hose connec-

tion of valve head.

Slide on the tube and the hose clip (U) until

touching the nose at the valve head.

Close hose clip (U) with flat pliers.

Slide the blinds (I) in the head cover.

I

Advertising