Readjustment of vnc 2, Notice – VACUUBRAND PC 600 LAN NT User Manual

Page 58

Advertising
background image

page 58 of 102

The vacuum gauge was adjusted using factory standards,

which are traceable through regular calibration in an ac-

credited laboratory (DAkkS calibration laboratory) to the

German national pressure standard. Depending on the

process and/or accuracy requirements, check the adjust-

ment and readjust if necessary. For readjustment, the de-

vice has to be adjusted both at atmospheric pressure as

well as under vacuum but only if the reference pressures

are known with certainty. The adjustment mode can be ac-

tivated only if the process control is inactive. Press ”Esc”

key, if necessary. If no adjustment is possible in the actual

pressure range:

”not possible” is displayed.

Adjustment at atmospheric pressure

An adjustment at atmospheric pressure is only

possible if the pressure is higher than 525 Torr

(700 mbar).

Vent the measurement connection of the VNC

2 and/or the vacuum system. Make sure that

the pressure transducer is at atmospheric

pressure.

In ”Device menu”, select program ”Cal.Sens” at the controller.

Use the ▲ and ▼ keys to adjust the reading to the current atmospher-

ic pressure.

Press the ”Enter” key to confirm.

1013 mbar

Note: To determine the actual atmospheric pressure, use an accurate

barometer or get accurate reading from the weather service, or a nearby

airport or other reliable source (taking into account the difference in alti-

tude between the source and the laboratory).

Readjustment of VNC 2