Cleaning the pressure transducer, See section “cleaning the pres, Sure transducer”, pg. 61 – VACUUBRAND PC 600 LAN NT User Manual

Page 61: Readjustment of the vnc 2 controller, Please continue with part 2 of this manual

Advertising
background image

page 61 of 102

The controller itself is maintenance-free.

Contamination of the pressure transducer or deposits will influence the

accuracy of measurement.

Attention: Never use a pointed or sharp-edged tool

to clean the pressure transducer.

Never touch the ceramic diaphragm of the pres-

sure transducer with hard objects.

Fill the measurement chamber with a solvent (e.g.,

benzene) and allow sufficient cleaning time. Observe

all regulations concerning usage and disposal of sol-

vents!

Drain the solvent and dispose of in accordance with

regulations. Repeat cleaning if necessary.

Rinse the measurement chamber several times with

alcohol in order to remove all solvent residues.

Allow the pressure transducer to dry.

Readjust the pressure transducer if necessary.

Readjustment of the VNC 2 controller

See section “Readjustment of VNC 2”, pg. 58.

Cleaning the pressure transducer

Please continue with part 2 of this manual.