Write commands compatible to ”cvc 3000, Write commands compatible to ” cvc 3000 – VACUUBRAND PC 600 LAN NT User Manual

Page 72

Advertising
background image

page 72 of 102

Command

Operation

Response

Description

IN_SP_5

switch-off

pressure

XXXX mbar/Torr/hPa

”Maximum” for ”vacuum control”,

”Minimum” for ”continuous pump-

ing”)

IN_SP_6

process runtime

XX:XX h:m

hours:minutes (00:00 = off)

Write commands compatible to CVC 3000

Command

Operation

Parameter Description

OUT_MODE

basic mode

0

1

2

4

VACUU•LAN

continuous pumping

vacuum control

RC 5 / RC 6 management

Attention: If control is running, only

switching from 1 to 2 and from 2 to 1 is

possible.

OUT-SP_1

set vacuum

XXXX

unit according to preselection; (0001 to

1060 mbar (hPa) or 0001 to 0795 Torr)

OUT_SP_V

set vacuum with

venting*

XXXX

unit according to preselection; (0001 to

1060 mbar (hPa) or 0001 to 0795 Torr)

OUT_SP_2

speed

XXX

speed in % (100% = ”HI”)

OUT_SP_3

start-up pressure

XXXX

unit according to preselection; (0001 to

1060 mbar (hPa) or 0001 to 0795 Torr)

OUT_SP_4

delay

XX:XX

hh:mm (hours:minutes)

OUT_SP_5

switch-off

pressure

XXXX

unit according to preselection; (0001 to

1060 mbar (hPa) or 0001 to 0795 Torr)

OUT_SP_6

time for automatic

switching off

(VACUU•LAN)

XX:XX

hh:mm (hours:minutes)

(00:00 = switch-off)

START

started

STOP

1

2

Stop

Stop and adopt the set vacuum

REMOTE**

0

1

Remote off

Remote on

CVC***

2

3

CVC 2000 commands

CVC 3000 commands

OUT_VENT

0

1

venting valve closed

venting valve open

STORE

store settings permanently

Advertising