VACUUBRAND PC 600 LAN NT User Manual

Page 69

Advertising
background image

page 69 of 102

* Pressure setting with venting is only possible in ”Vacuum control”

mode if a venting valve is connected to the VNC 2 and configured,

and vacuum control is started. The venting valve opens automatically

if the actual pressure is at least 7.5 Torr (10 mbar) below the pre-

set pressure. Automatic venting becomes inactive if vacuum control

is stopped (by pressing ”Esc”), a pressure value is set using the com-

mand OUT_SP_1, or if the basic mode is changed. Activate the com-

mand OUT_SP_V again if necessary.

** If remote operation is selected or deselected, the user has to ensure

that no dangerous status of the system can occur due to the change of

the mode of operation, and must take appropriate safety precautions,

especially if selecting remote operation interferes with a locally oper-

ated active process.

Command

Operation

Parameter

Description

OUT_VENT

driving venting

valve

0

1

close venting valve (not automatically)

open venting valve (process control

stopped, valve does not close again

automatically)

STORE

store settings

permanently