Fault

Possible cause

Remedy

❑ Controller does

not respond when

pressing keys (ex-

cept ON/OFF). ”P”

is displayed.

➨ Controller in remote

mode?

✔ Control VNC 2 controller

via interface or switch off

remote mode.

❑ Controller does

not respond when

operating any keys.

No change after

switching off/on.

✔ Contact local distributor.

❑ Pump does not

work.

➨ Pump too hot?

✔ Allow pump to cool down,

identify and eliminate

cause of failure. Manual

reset is necessary. Switch

off pump or unplug.

➨ Electrical power cord

not plugged in, electri-

cal supply failure?

✔ Plug in power cord.

Check fuse.

❑ Pump does not

start or stops im-

mediately.

➨ Motor overloaded?

✔ Allow motor to cool down,

identify and eliminate

cause of failure. Manual

reset is necessary. Switch

off pump or unplug.

➨ Overpressure in outlet

line?

✔ Remove blockage in line,

open valve.

➨ Pump has been ex-

posed to condensate?

✔ Perform maintenance.

❑ Pump does not

achieve its ultimate

vacuum or usual

pumping speed.

➨ Outgassing substances

or vapor generated in

the process?

✔ Check process parame-

ters.

➨ Leak in the tubing or in

the vacuum system?

✔ Check connections, pipe-

line and vacuum system.

➨ Wrong setting at con-

troller?

✔ Select mode ”continu-

ous pumping” and check

again.

➨ Long, narrow vacuum

line?

✔ Use lines with larger di-

ameter, length as short as

possible.

➨ Pump has been ex-

posed to condensate?

✔ Allow pump to run for

some minutes with atmo-

spheric pressure at the

inlet to purge.