Notes on operating the controller – VACUUBRAND PC 600 LAN NT User Manual

Page 44

Advertising
background image

page 44 of 102

Notes on operating the controller

Valves and/or vacuum pumps are necessary to operate the control-

ler.
Note:
Texts written in Courier font mirror the display of the

controller.

+

Inside a vacuum system where evaporation

occurs, e.g., a rotary evaporator, the vacu-

um is not uniform. For example, a condens-

er can act as a pump, or the vacuum in the

connecting tubing can be higher or lower

than in the application itself. This affects the

measurement results as well as the control

levels.

Therefore, carefully choose the position where to con-

nect an external (optional) gauge head. Position an

optional external gauge head and its vacuum line in

such a way that condensate cannot flow towards the

pressure transducer. Condensate and deposits will

affect the measurement results. Clean the pressure

transducer, if necessary. See section ”Cleaning the

pressure transducer”, pg. 61.

vacuum

connection

+

Maximum permissible pressure: 21.8 psi (1.5 bar) ab-

solute.

+

Attention: At pressures above approximately 810 Torr

(1080 mbar) the pressure reading becomes incorrect

due to saturation of the pressure transducer. The dis-

play will flash. Release pressure immediately! Risk of

bursting!

+

Setting of the interface, see

“Interface parameters”, pg. 65.

+

Presettings at the controller, see “General view of factory-set basic

modes”, pg. 45 and “Menu guide”, pg. 47.

+

Operating the controller, see “Working with the controller”, pg. 52.