Adjustment under vacuum, Adjustment at a reference pressure – VACUUBRAND PC 600 LAN NT User Manual

Page 59

Advertising
background image

page 59 of 102

Adjustment under vacuum

An adjustment under vacuum is only possible

if the pressure is lower than 15 Torr (20 mbar)

absolute.

Evacuate the measurement connection of the

VNC 2 to a pressure < 0.5 Torr (mbar) (e.g. by

applying a good two-stage rotary vane pump).

In ”Device menu”, select program ”Cal.Sens” at the controller.

+

The reading is automatically adjusted to ”zero”.

Press the ”Enter” key to confirm.

0 mbar

Note: Adjustment under vacuum with an actual pressure higher than 0.5

Torr (mbar) reduces the accuracy of the measurement. If the pressure is

significantly higher than 0.5 Torr (mbar), adjustment to a reference pres-

sure is recommended.

Adjustment at a reference pressure

Instead of adjustment under vacuum to a pres-

sure < 0.5 Torr (mbar), adjustment to a precise-

ly known reference pressure within the range

of 0..... 15 Torr (20 mbar) is possible. Evacuate

the measurement connection of the VNC 2 to

a pressure within 0 ..... 15 Torr (0.....20 mbar).

In ”Device menu”, select program ”Cal.Sens” at the controller.

+

The reading is automatically adjusted to ”zero”.

Use the ▲ and ▼ keys to adjust the reading to the reference pressure

at the vacuum line within the range of 0 ..... 15 Torr (0.....20 mbar).

Press the ”Enter” key to confirm.

0...20 mbar

Note: The accuracy of the value of the reference pressure will directly

affect the accuracy of the adjustment. If the nominal ultimate vacuum

of a diaphragm pump is used as reference vacuum, the accuracy of the

controller might be doubtful. The diaphragm pump may not achieve the

specified value (due to condensate, poor condition, failure of valves or

diaphragm, leaks).