Troubleshooting – VACUUBRAND PC 600 LAN NT User Manual

Page 77

Advertising
background image

page 77 of 102

Fault

Possible cause

Remedy

❑ No display.

➨ Electrical power cord

not plugged in, electri-

cal supply failure?

✔ Plug in power cord.

Check fuse.

➨ Device fuse blown?

✔ Identify cause of failure.

Replace device fuse at

switch output of VNC 2 or

at diaphragm pump.

➨ Device switched off?

✔ Switch on controller and/

or pump.

➨ Other than above men-

tioned causes?

✔ Contact local distributor.

❑ Display disappears,

internal overload

protection activated.

➨ Too much load (e.g.,

valves) connected?

✔ Check current draw of

the connected devices

(pump, valves).

➨ Thermal overload at

controller, ambient tem-

perature too high?

✔ Ensure sufficient ventila-

tion.

➨ Short circuit at connect-

ed valves?

✔ Replace valves.

➨ Short circuit at the

RS 232 plug?

✔ Check plug and cable.

➨ Other than above men-

tioned causes?

✔ Contact local distributor.

❑ Pressure reading

incorrect.

➨ Pressure transducer

decalibrated?

✔ Readjust VNC 2.

➨ Humidity in the mea-

surement chamber?

✔ Let the pressure trans-

ducer dry, e.g., by pump-

ing. Readjust if neces-

sary. Determine and

eliminate the cause for

humidity.

➨ Pressure transducer

contaminated?

✔ See ”Cleaning the pres-

sure transducer”.

➨ Other than above men-

tioned causes?

✔ Contact local distributor.

❑ Error message

”ErrValve”, two

blips*.

➨ External venting valve

removed or defective?

✔ Connect or check valve or

replace with a new one or

reconfigure without valve.

Switch VNC 2 off/on.

Troubleshooting