Shutdown & storage, Notice – VACUUBRAND PC 600 LAN NT User Manual

Page 40

Advertising
background image

page 40 of 102

Permissible range of coolant temperature at the exhaust

waste vapor condenser:

5 °F to 68 °F (-15°C to +20°C)

Check hose connections prior to starting operation of the

cooling system.

Check coolant hoses regularly during operation.

Removing the catchpots:

Catchpot at outlet:

Remove joint clip. Remove catchpot

and drain conden-

sate.

Catchpot at inlet:

Admit air or inert gas (via the pump inlet) to restore at-

mospheric pressure in the catchpot before attempting

removal. Remove joint clip. Remove catchpot and drain

condensate.

Reattach drained catchpots.

+

Important: Comply with regulations when disposing

of solvents/condensates. Recycle if possible; purify if

contaminated.

Shutdown & storage

The pump can be switched off under vacuum.

Short-term:

Has the pump been exposed to condensate?

- Allow the pump to continue to run at atmospheric pres-

sure for a few minutes.

Has the pump been exposed to media which may damage

the pump materials or form

deposits?

- Check and clean pump heads if necessary.

Has the pressure transducer been exposed to media which

may form deposits?

NOTICE

NOTICE