Important notes concerning the operation, Of the exhaust waste vapor condenser, Notice – VACUUBRAND PC 600 LAN NT User Manual

Page 39

Advertising
background image

page 39 of 102

Important notes concerning the operation of

the exhaust waste vapor condenser

Connect the exhaust to a suitable treatment system to

prevent the discharge of dangerous gases and vapors

to the surrounding atmosphere.

+

Never block the gas outlet ((2) hose nozzle for tubing

I.D. 3/8” (10 mm)). The exhaust hose must always be

unobstructed and without back-pressure to enable an

unhindered discharge of gases.

+

Check the overpressure safety relief device (11) at the

exhaust waste vapor condenser (10) regularly; replace

if necessary. Check especially for deterioration, co-

alescence and cracks.

• Ensure that the

coolant outlet hose is always free

and that it cannot get blocked.

• Maximum permissible coolant pressure at the exhaust

waste vapor condenser: 87 psi (6 bar) absolute

• Comply with the maximum permissible coolant pres-

sures of additional components in the coolant circuit

(e.g., coolant valve).

• We strongly recommend installing an optional coolant

valve (see “Accessories”, pg. 74)

in the supply line

of the exhaust vapor condenser to save water and re-

duce the risk of water spill.

• Avoid overpressure in the coolant circuit (e.g., caused

by blocked or kinked coolant hoses).

In case of

condensation: Check the liquid level in both

catchpots (9) during operation. Check the liquid level reg-

ularly. Do not allow the catchpots to overfill. Drain catch-

pots in time to avoid overflow. Install a level sensor (see

“Accessories”, pg. 74) for monitoring, if necessary.

The maximum liquid level is at approximately 80% of the

total filling level to avoid problems when removing the

catchpots.