Write commands compatible to ”cvc 2000, Write commands compatible to ” cvc 2000 – VACUUBRAND PC 600 LAN NT User Manual

Page 68

Advertising
background image

page 68 of 102

Write commands compatible to CVC 2000

Command

Operation

Parameter

Description

OUT_MODE

basic mode

0

1

2

4

VACUU•LAN

continuous pumping

vacuum control

RC 5 / RC 6 management

Attention: If control is running, only

switching from 1 to 2 and from 2 to 1 is

possible.

OUT_SP_1

set vacuum

XXXX

unit according to preselection; (0001 to

1060 mbar (hPa) or 0001 to 0795 Torr)

OUT_SP_V

set vacuum

with venting*

XXXX

unit according to preselection; (0001 to

1060 mbar (hPa) or 0001 to 0795 Torr)

OUT_SP_2

set frequency

XX.X

motor speed in Hz (01.0 to 60.0 in steps

of 0.5 Hz or 99.9 for

“HI”

)

OUT_SP_3

vacuum for restart

(VACUU•LAN)

XXXX

unit according to preselection; (0001 to

1060 mbar (hPa) or 0001 to 0795 Torr)

OUT_SP_4

delay

XX:XX

hh:mm (hours:minutes)

OUT_SP_5

vacuum for auto-

matic switching

off

XXXX

unit according to preselection; (0001 to

1060 mbar (hPa) or 0001 to 0795 Torr)

OUT_SP_6

time for automatic

switching off

(VACUU•LAN)

XX:XX

hh:mm (hours:minutes)

(00:00 = switch-off)

START

starting process

control

STOP

stopping process

control

1

2

termination of process control

termination of process control and stor-

age of the current pressure as new set

point

REMOTE

remote

operation**

0

1

set controller to local operation

set controller to remote operation

Command

Operation

Response

Description

IN_STAT

status of

process control

XX40

XX41

XX42

XX43

RC 5 / RC 6 management: not active

RC 5 / RC 6 management:

pumping down

RC 5 / RC 6 management:

preset vacuum is achieved

RC 5 / RC 6 management:

system is switched off

Advertising
This manual is related to the following products: