Connection of components at the controller

At the side of the controller are the serial in-

terface RS-232 C and VACUU•BUS jacks for

connection of accessory components: in-line

valve / coolant valve / external pressure trans-

ducer / external venting valve / etc.
+

Plug the VACUU•BUS lines of accessory

components into the VACUU•BUS jacks. Do

not apply off-axis forces when assembling

or removing plug connections! Observe cor-

rect orientation of the plugs.

Catchpots:

The catchpot at the inlet protects against drop-

lets and particles from entering the pump.
+

Enhances lifetimes of diaphragms and

valves.

+

Improves vacuum performance in applica-

tions with condensable vapors.

Both catchpots are coated with a protective

layer to protect against shattering in case of

breakage or implosion.

Assemble the catchpots at the inlet and at

the outlet using joint clips.

Exhaust waste vapor condenser:

Assemble the hose nozzles for coolant inlet

(12) and coolant outlet (13) tubing at the ex-

haust waste vapor condenser.

The exhaust waste vapor condenser enables

an efficient condensation of the pumped va-

pors at the outlet.

+

No backflow of condensates.

+

Controlled recovery of condensates.

+

Close to 100% solvent recovery.

+

The isolation cover protects against glass splinters in

case of breakage, acts as thermal isolation to avoid

catchpot at

the outlet

catchpot at

the inlet

(9)

overpressure

safety relief

device

outlet (gas!)

2

12

13