VACUUBRAND PC 600 LAN NT User Manual

Page 73

Advertising
background image

page 73 of 102

* Pressure setting with venting is only possible in ”Vacuum control” mode

if a venting valve is connected and configured, and vacuum control is

started. The venting valve opens automatically if the actual pressure is

at least 7.5 Torr (10 mbar) below the preset pressure. Automatic vent-

ing becomes inactive if vacuum control is stopped (by pressing ”Esc”),

a pressure value is set using the command OUT_SP_1, or if the basic

mode is changed. Activate the command OUT_SP_V again if neces-

sary.

** If remote operation is selected or deselected, the user has to ensure

that no dangerous status of the system can occur due to the change of

the mode of operation, and must also take appropriate safety precau-

tions, especially if selecting remote operation interferes with a locally

operated active process.

*** After being switched on, the controller is in ”CVC 2” mode by default.

Send ”CVC 3” and ”STORE” to permanently set the controller RS 232C

commands to the extended set ”CVC 3000”.