Notice – VACUUBRAND PC 600 LAN NT User Manual

Page 82

Advertising
background image

page 82 of 102

Tools required (metric):
- Torx driver T20

- 5 mm wide Allen key

- 2.5 mm wide slotted screwdriver

- Flat pliers

- Diaphragm key width 66 mm

Service kit MZ 2C NT for PC 500 LAN NT...................................................696869

Service kit MD 4C NT for PC 600 LAN NT ..................................................696870

Diaphragm key (width 66 mm) ...................................................................636554

Ensure that maintenance is done only by suitably

trained and supervised technicians.

The valves and diaphragms as well as the motor capac-

itors are wear parts. If the rated ultimate vacuum is no

longer achieved or in case of increased noise level, the

pump interior, the diaphragms and the valves must be

cleaned and the diaphragms and valves must be checked

for cracks or other damage.

All bearings are encapsulated and are filled with long-life

lubricant. Under normal operating conditions, the drive

system is maintenance free.

In demanding circumstances, it may be efficient to check

and clean the pump heads on a regular basis. In normal

use, the lifetime of the diaphragms and valves is typically

15,000 operating hours.

- Prevent internal condensation, transfer of liquids or

dust. The diaphragms and valves will be damaged if

liquid is pumped in significant amount.

- Carry out maintenance frequently if the pump is ex-

posed to corrosive media or in case of deposits.

- Regular maintenance will improve the lifetime of the

pump and also protect both users and the environment.