VACUUBRAND PC 600 LAN NT User Manual

Page 26

Advertising
background image

page 26 of 102

Type

PC 500 LAN NT PC 600 LAN NT

Degree of protection IEC 529

IP 40

Inlet

connection for PTFE tube

O.D. 3/8”, I.D. 5/16” (10 / 8 mm)

Outlet

hose nozzle for tubing I.D. 3/8”

(hose nozzle DN 10 mm)

Coolant connection

(waste vapor condenser)

hose nozzle for tubing I.D. 1/4” - 5/16”

(hose nozzle DN 6-8 mm)

Maximum permissible

pressure of coolant at

waste vapor condenser

psi

(bar)

87 (absolute)

(6 (absolute))

Permissible range of cool-

ant temperature (waste

vapor condenser)

°F

(°C)

5 to 68

(-15 to +20)

Volume of catchpot

(only pumping unit)

qt

(ml)

0.52

(500)

A-weighted emission

sound pressure level**

(uncertainty K

pA

: 3 dB(A))

dB(A)

45

Dimensions L x W x H

approx.

in

(mm)

16.1 x 10.4 x 15.8

(408 x 264 x 402)

16.1 x 10.4 x 18.5

(408 x 264 x 470)

Weight approx.

lbs. (kg)

33.7 (15.3)

40.6 (18.4)

** Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 230V/50Hz and ulti-

mate vacuum with exhaust tube at outlet.

Controller

VNC 2

Pressure transducer

ceramic diaphragm (alumina),

capacitive, absolute pressure,

gas type independent

Display

LCD alphanumeric display, illuminated

2 x 8 characters

Pressure units / scale (selectable)

Torr, mbar or hPa

Measuring range (absolute)

825 - 1 Torr (1100 - 1 mbar)

Maximum control range (absolute)*

795 - 1 Torr (1060 - 1 mbar)

Maximum permissible pressure at pres-

sure transducer (absolute)

1125 Torr (1.5 bar)

Maximum permissible temperature of

gaseous media**

continuous operation: 104°F (40°C),

for short periods (less than 5 minutes)

up to 176°F (80°C)

Advertising