Interface parameters

The VNC 2 controller is equipped with a serial interface (RS 232C, nine-

pin Sub-D-plug).
+

Plug-in or remove the cable (cable RS 232C) from the interface only if

the equipment is switched off.

+

The interface is

not electrically isolated from the measuring circuit.

+

For optimal electromagnetic compatibility assemble an interface filter

(cat. no.: 638235).

The controller is fully operable via the serial interface. Measuring results,

preselections and the status of the controller can be read at any time.

The factory-set read and write commands of the VNC 2 are completely

compatible with the VACUUBRAND CVC 2000 controller (see sections

”Read / Write commands compatible to CVC 2000”). An extended instruc-

tion set for the VNC 2 controller is available using the command ”CVC 3”

(see sections ”Read / Write commands compatible to CVC 3000”).