4 location of installation and ambient conditions, Caution – BINDER MK 115 User Manual

Page 29

Advertising
background image

MKF + MKFT (E3.1) 07/2014

page 29/114


With temperatures below +3 °C / 37.4 °F, water must be completely removed from the humidifying sys-

tem.

CAUTION

Storage below +3 °C / 37.4 °F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the unit.

Contact BINDER Service before any storage below +3 °C / 37.4 °F.

Permissible ambient humidity: max. 70 % r.H., non-condensing

CAUTION

Condensation by excess humidity.
Danger of corrosion on the housing after operating at humidity values > 70 % r.H.

for a long period.

Dry the appliance completely before shut-down:

Set the humidity to 0 % r.H. or turn off humidity switch (4).

Set the temperature set point to 150 °C / 302 °F (Manual mode). Let the unit oper-

ate for approx. 2-3 hours with closed door. Remove the access port plugs.

Only then, shut down the unit at the main power switch (3) and close the tap of the

water supply.

After drying the unit for decommissioning, the humidity value will approximate ambient humidi-

ty.

When after storage in a cold location you transfer the unit to its warmer installation site, condensation

may form. Before start-up, wait at least two hours until the chamber has attained ambient temperature

and is completely dry and the oil in the compressors has warmed up.
In case of a prolonged temporal decommissioning: Leave the unit door open or remove the access port

plugs.

3.4 Location of installation and ambient conditions

Set up the alternating climate chamber on a flat, even and non-flammable surface, free from vibration,

and in a well-ventilated, dry location and align it using a spirit level. The site of installation must be capa-

ble of supporting the unit’s weight (see technical data, chap. 20.4). The chambers are designed for setting

up inside a building (indoor use).
When after storage in a cold location you transfer the unit to its warmer installation site, condensation

may form. Before start-up, wait at least two hours until the chamber has attained ambient temperature

and is completely dry and the oil in the compressors has warmed up.

CAUTION

Danger of overheating.
Damage to the unit.

Do NOT set up units in non-ventilated recesses.

Ensure sufficient ventilation for dispersal of the heat.