MKF + MKFT (E3.1) 07/2014

page 70/114


Depending on size, material, and shape of the charging material and on the heating-up rate, condensa-

tion may form despite the activated bedew protection. This condensation is, however, reduced compared

to the state without bedew protection.

11. Zero-voltage relay outputs via operation lines 2 to 5

Operation lines 2 to 5 serve to switch any device connected to the zero-voltage relay output (DIN sockets

(11) and (12) located in the lateral control panel). They can be programmed ON/OFF in Manual Mode

(chap. 8) as well as in Program Mode (AUTO, chap. 9) via operation lines 2 to 5.
Connection for operation lines 2 and 3 occurs via DIN socket (11), connection for operation lines 4 and 5

via DIN socket (12) in the lateral control panel:

OUTPUT TRACK 2+3

24V/MAX.2,5A

OUTPUT TRACK 4+5

24V/MAX.2,5A

Figure 18: Pin configuration of DIN sockets (11) left and (12) right



DIN socket (11):

Operation line 2

Operation line 3

Pin 1: Pin
Pin 2: Make

Pin 4: Pin
Pin 5: Make


DIN socket (12):

Operation line 4

Operation line 5

Pin 1: Pin
Pin 2: Make

Pin 4: Pin
Pin 5: Make


Maximum loading capacity of the switching contacts: 24V AC/DC – 2.5 A

DANGER

Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.

Do NOT exceed the maximum switching load of 24V AC/DC – 2.5A.

Do NOT connect any devices with a higher loading capacity.