MKF + MKFT (E3.1) 07/2014

page 63/114

9.7 Repetition of a section or several sections within a program

Here we use the example of a set-point ramp temperature program of chap. 9.5. The shaded sections 02

and 03 shall be repeated e.g. 30 times.

Program

section

No.

Set-

point

temp.

W-1

Fan

FAN

Section time

Time

Operation lines

Sk

Target

section

No

No. of

cycles

Cy

Minimum

tolerance

Tmin

Maximum

tolerance

Tmax

01

-20

****.

00:30:00

00000000

1

0

-1999

+9999

02

0

****.

01:30:00

00000000

1

0

-5

+5

03

60

****.

01:00:00

00000000

1

0

-2

+2

04

60

****.

03:20:00

00000000

1

0

-1999

+9999

05

-40

****.

00:00:01

00000000

1

0

-1999

+9999

The following table shows the program that results, whereby the differences to the table above are shad-

ed.

Program

section

No.

Set-

point

temp.

W-1

Fan

FAN

Section time

Time

Operation lines

Sk

Target

section

No

No. of

cycles

Cy

Minimum

tolerance

Tmin

Maximum

tolerance

Tmax

01

-20

****.

00:30:00

00000000

1

0

-1999

+9999

02

0

****.

01:30:00

00000000

1

0

-5

+5

03

60

****.

01:00:00

00000000

2

30

-2

+2

04

60

****.

03:20:00

00000000

1

0

-1999

+9999

05

-40

****.

00:00:01

00000000

1

0

-1999

+9999

Sections 02 and 03 will be executed in total 31 times; only then will the program continue.
Entry of the values into the display program table:

To have sections repeated indefinitely, enter the number of cycles “Cy” as -1.

9.8 Performance after power failure in Program Mode

The program is resumed at the point where the interruption occurred with the latest set-points reached

during the program run. The power failure is noted in the event list. No error message is displayed indicat-

ing that a power failure had taken place.