BINDER MK 115 User Manual

Page 52

Advertising
background image

MKF + MKFT (E3.1) 07/2014

page 52/114


Ranges:

Temperature

MKF

Setting range (range -50 °C up to -40 °C

not provided for operation)

-50 °C / -58°F up to 180 °C / 356°F

Control range without humidity

- 40ºC / -40°F to + 180ºC / 356°F

Control range in climatic operation

+ 10 °C / 50°F up to +95 °C / 203°F

Control range in climatic operation, with

optional compressed air dryer

0 °C / 32 °F up to +95 °C / 203°F

Temperature

MKFT

Setting range (range -80 °C up to -70 °C

not provided for operation)

-80 °C / -112 °F up to 180 °C / 356°F

Control range without humidity

-70 ºC / -94 °F up to 180 °C / 356°F

Control range in climatic operation

+ 10 °C / 50°F up to +95 °C / 203°F

Control range in climatic operation, with

optional compressed air dryer

0 °C / 32 °F up to +95 °C / 203°F

Humidity

Setting range

0 % r.H. up to 100 % r.H.

Control range

10 % r.H. up to 98 % r.H.

Control range with optional compressed

air dryer

5 % r.H. up to 98% r.H.

For possible combinations of temperature and humidity values without condensation,

see temperature / humidity diagram in chap. 14. Outside the indicated control range for

temperature and humidity the humidity system is automatically turned off. The entry of

the humidity set-point 0 % r.H. permits completely turning off humidity in defined pro-

gram section and thus attaining faster temperature changes.

For the control range of temperature and relative humidity, see the temperature / humidity diagram (chap.

14).

With set-point type “

Limit”, adapt the safety controller (chap. 12.2) always when you changed

the temperature set-point. Set the safety controller set-point by approx. 10 °C above the con-

troller temperature set-point.
We recommend keyboard locking (option, chap. 16.5.) during operation.

In case of the optional temperature safety device (chap. 12.3), check and, if necessary, adjust

the temperature limits entered there.

In Manual Mode, no program can be started. Set-points can be entered for temperature and for humidity.

The actual values equilibrate to these set-points.
When pushing the EXIT button in Manual Mode, the controller changes to Idle Mode. The set-points en-

tered in Manual Mode remain saved.

When incidentally pressing the EXIT button during Manual Mode operation, the controller will

change to Idle Mode and thus will not adjust any longer to the program set-points.
We recommend keyboard locking (option, chap. 16.5.) during operation.

For a negative set-point entry, enter the numerical value first and then the minus sign (-).

When operating without humidity (humidity switch (4) OFF), set the humidity set-point in Manu-

al Mode to 0 % r.H. in order to avoid alarms (in case of the humidity deviating by more than +/-

5% from the set-point).

Advertising
This manual is related to the following products: