19. Troubleshooting

Fault description

Possible cause

Required measures

Heating
Chamber without function.

Turning on the main power

switch (3) has no effect

Rear power switch (20) not turned on.

Turn on the rear power switch

(20) at least one hour before

operating the chamber.

Chamber heating permanently,

set-point not maintained.

Semiconductor relay defective.

Contact BINDER service.

Pt 100 sensor defective
Controller defective.
Controller not adjusted.

Calibrate and adjust controller.

Chamber doesn’t heat up.

Heating element defective.

Contact BINDER service.

Semiconductor relay defective

Chamber doesn’t heat up when

turned on.

Safety controller responds.

Limit temperature reached. Safety

controller (chap. 12.2) set too low.

Let the chamber cool down and

press the RESET button of the

MB1 controller. If appropriate,

select suitable limit value.

Safety controller (chap. 12.2) defec-

tive.

Contact BINDER service.

Unit permanently turned off.

Nominal temperature exceeded by 20

°C due to unit failure. Over tempera-

ture protective device (class 1) re-

sponds.

Contact BINDER service.

Safety device class 2 re-

sponds.

Limit temperature reached.

Disconnect the chamber from

the power supply and let it cool

down. Detect the cause and

remove it. Press the “RESET”

button at the controller.

Start up the chamber and check

control functions. If appropriate,

select suitable limit value.

Over-/under temperature safety

device class 2 (option) re-

sponds.

Limit temperature reached.

Disconnect the chamber from

the power supply and let it cool

down. Detect cause and re-

move it. Press button “RESET

CL 2.0” (5).

Start up the chamber and check

control functions. If appropriate,

select suitable limit value.

Refrigerating performance

No or low refrigerating perfor-

mance.

Ambient temperature > 25 °C / 77°F

(chap.3.4).

Select cooler place of installa-

tion.

Combination of temperature/humidity

values not in the optimum range (see

temperature humidity diagram, Figure

19).

Select combination of tempera-

ture/humidity values in the op-

timum range (chap. 14).

Compressor not turned on.

Contact BINDER service.

Electro-valves defective.
No or not enough refrigerant.

No refrigerating performance;

notification “1H PREHEAT

PHASE” in the controller dis-

play.

Rear power switch (20) turned on

less than 1 hour before operating the

chamber.

Turn on the rear power switch

(20) at least one hour before

operating the chamber.