MKF + MKFT (E3.1) 07/2014

page 35/114

4.3 Connection of cooling water outlet for water cooling (option)

An enclosure inside the unit contains the connection kit for the cooling water inlet and outlet.

Fasten the cooling hose to the connection “OUT” (21) (Figure 8) on the rear of the unit (screw thread

¾’’).

You can use a part of the supplied tap water hose as a drainage hose. In case another hose is used, it

has to be permanently resistant against max. 50 °C / 122 °F.

Protect both ends of the drainage hose with two of the four supplied hose clamps. Before turning on

the unit, check the connection for leaks.


4.4 Connection of cooling water inlet for water cooling (option)

Connect the cooling water outlet

before connecting the cooling water inlet.

Type of suitable water quality:

Water intake temperature: max. 10 °C / 50 °F.

pH value 4-7

connection pressure: 4 to 10 bar

BINDER GmbH is NOT responsible for the water quality at the user’s site.
Any problems and malfunctions that might arise following use of water of deviating quality is

excluded from liability by BINDER GmbH.
The warranty becomes void in the event of use of water of deviating quality.

An enclosure inside the unit contains the connection kit for the cooling water inlet and outlet.

Fasten the cooling water hose to the connection “IN” (22) (Figure 8) on the rear of the unit (screw

thread ¾’’).

Install the water supply connection using either the enclosed water hose or another pressure-resistant

one. To accomplish this, remove the cover of the freshwater connection “IN” (22) (Figure 8) on the rear

of the unit.

Protect both ends of the hose with two of the four supplied hose clamps. Before turning on the unit,

check the connection for leaks.