MKF + MKFT (E3.1) 07/2014

page 94/114


Fault description

Possible cause

Required measures

Controller (continued)
Chart recorder function: meas-

ured-value memory cleared,

information lost.

New setting of storage rate.

Change the storage rate ONLY

if the previously registered data

are no longer required (chap.

7).

Controller does not attain set-

points entered in Manual

Mode.

Button EXIT or AUTOMATIC has

been pressed: Unit is in Idle Mode.

Change to Manual Mode (chap.

8).

Controller does not attain pro-

gram set-points.

Button EXIT or AUTOMATIC has

been pressed: Unit is in Idle Mode.

Start the program again (chap.

9.9).

Program duration longer than

programmed.

Tolerances have been programmed.

For rapid transition phases, do

NOT program tolerance limits in

order to permit maximum heat-

ing, refrigerating, or humidifica-

tion speed.

Program stops one section too

early.

Program line is incomplete.

When programming, define the

end value of the desired cycle

by adding an additional section

with a section time of at least

one second.

Ramp temperature transitions

are only realized as steps.

When using the Program Editor of the

software APT-COM™ 3 DataControl-

System, the setting “step” has been

selected.

Select setting “ramp” in the

Program Editor of the software

APT-COM™ 3 DataControlSys-

tem and transfer a program to

the chamber controller.

Humidity alarm message when

operating without humidity

(humidity switch (4) OFF)

Humidity set-point set to a value

> 0% r.H.

Manual Mode: Enter a humidity

set-point 0% r.H.

Program Mode: Enter a humidi-

ty subprogram with humidity

set-point 0% r.H.

Display flashing:

1999 or -1999 or 9999.

Sensor rupture between sensor and

controller or Pt 100 sensor defective. Contact BINDER service.

Short-circuit.

Initialization problem due to turning

on the chamber too early.

Observe a delay time of approx.

30s between turning the cham-

ber Off and On again the

chamber.

Only qualified service personnel authorized by BINDER must perform repair.

Repaired units must comply with the BINDER quality standards.