Ole_link2, Ole_link3, Ole_link4 – BINDER MK 115 User Manual

Page 97: Ole_link5

Advertising
background image

MKF + MKFT (E3.1) 07/2014

page 97/114

Unit size

115

240

720

Electrical data
IP-system of protection acc. to EN 60529

IP

20

20

20

Nominal voltage (+/-10%) 50 Hz

V

400 3N~

400 3N~

400 3N~

Nominal power

kW

4.50

6.80

11.00

Energy consumption

at +25 °C / 77 °F and 60 % r.H.

Wh/h

1250

1500

3900

Power plug: CEE plug 5-poles

Amp

16

16

32

Over-voltage category acc.

to

IEC 61010-1

II

II

II

Pollution degree acc. to IEC 61010-1

2

2

2

Over-current release category B,

3 x internal

Amp

16

16

25

Electrical data of the voltage and frequency changer

IP-system of protection acc. to EN 60529

IP

23

23

23

Nominal voltage (+/-10%) 60 Hz (input side)

V

480 3N~

480 3N~

480 3N~

Nominal power

kW

9

9

13

Over-voltage category acc.

to

IEC 61010-1

II

II

II

Pollution degree acc. to IEC 61010-1

2

2

2

Fuse

A

16

16

25

Further information
Weight (empty)

kg / lbs.

280 / 617

360 / 794 590 / 1300

Weight (empty) with optional compressed air dryer

kg / lbs.

295 / 650

375 / 827 605 / 1334

Filling weight of refrigerant R 404A (GWP 3750)

kg / lbs.

2.00 / 4.41 2.20 / 4.85 5.00 / 11.02

Noise level

approx. dB(A)

62

65

65

Noise level with optional compressed air dryer

(short-term)

approx. dB(A)

85

85

85

Noise level with voltage and frequency changer

approx. dB(A)

67

67

67

1)

Lower values are valid at an ambient temperature of max. 25 °C / 77°F

2)

Depending on the set-point

3)

Units with voltage and frequency changer: Values reduced by 0.3 K/min each.

*

Data not yet determined

All technical data is specified for unloaded units with standard equipment at an ambient temperature of

+25 °C / 77°F and a power supply voltage fluctuation of +/-10%. The temperature data is determined in

accordance to BINDER factory standard following DIN 12880, observing the recommended wall clear-

ances of 10 % of the height, width and depth of the inner chamber.

All indications are average values, typical for units produced in series. We reserve the right to

change technical specifications at any time.

If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-

cording to the load.

Bringing a source of humidity into the inner chamber will affect the minimum humidity specifi-

cation and may affect the humidity accuracy.

Advertising
This manual is related to the following products: