20.4 MKF (E3.1) technical data

Unit size

115

240

720

Exterior dimensions
Width (including 18 mm / 0.7 in for 1 access port

(MKF 115, 240), 36 mm / 1.4 in for 2 access ports

(MKF 720), with plug)

mm / inch 1000 / 39.37 1135/ 44.69 1615 / 63.58

Height (incl. castors)

mm / inch 1725 / 67.91 1715 / 67.52 2005 / 78.94

Depth (incl. cable and door handle)

mm / inch 915 / 36.02 1000/ 39.37 1230 / 48.43

Depth (incl. cable and door handle) with optional

compressed air dryer

mm / inch 1085 / 42.72 1170 / 46.06 1400 / 55.12

Depth (incl. cable and door handle) with voltage

and frequency changer

mm / inch 1530 / 60.24 1615 / 36.58 1845 / 72.64

Wall clearance rear

mm / inch 300 / 11.81 300 / 11.81 300 / 11.81

Wall clearance rear with optional compressed air

dryer or to set up the voltage and frequency

changer

mm / inch 1000 / 39.37 1000 / 39.37 1000 / 39.37

Wall clearance sides

mm / inch

200 / 7.87 200 / 7.87 200 / 7.87

Window width

mm / inch 288 / 11.34 508 / 19.99 508 / 19.99

Window height

mm / inch

222 / 8.74 300 / 11.81 300 / 11.81

Number of doors

pc.

1

1

1

Interior dimensions
Width

mm / inch 600 / 23.62 735 / 28.94 1200 / 47.24

Height

mm / inch 480 / 18.90 700 / 27.56 1020 / 40.16

Depth

mm / inch 400 / 15.75 443 / 17.44 600 / 23.62

Interior volume

l / cu.ft.

115 / 4.06 228 / 8.05 734 / 25.92

Number of racks, standard / max.

pc.

1/4

1/6

1/11

Load per rack

kg / lbs.

30 / 66

30 / 66

40 / 88

Permitted total load

kg / lbs.

60 / 132

70 / 155

160 / 353

Temperature data (without humidity)
Temperature range 1)

°C / °F

-40 to +180 / -40 to 356

Temperature fluctuation 2)

± K

0.1 to 0.6

0.1 to 0.5

0.1 to 0.5

Temperature uniformity (variation) 2)

± K

0.1 to 1.3

0.1 to 1.5

0.1 to 1.8

Average heating up time acc. to IEC 60068-3-5

3)

K/min.

5.5

5.0

4.8

Average cooling down time acc. to IEC 60068-3-5

K/min.

4.5

5.0

4.8

Max. heat compensation up to 25 °C / 77 °F

W

2500

2800

6500

Climatic data (with humidity)
Temperature range

°C / °F

+10 to +95 / 50 to 203

Temperature range with optional compressed air

dryer

°C / °F

0 to +95 / 32 to 203

Temperature fluctuation 2)

± K

0.1 to 1.3

0.1 to 1.3

0.2 to 1.5

Humidity range

% r.H.

10 to 98

10 to 98

10 to 98

Humidity range with optional compressed air dryer

% r.H.

5 to 98

5 to 98

5 to 98

Humidity fluctuation 2)

± % r.H.

≤ 2,5

≤ 2,5

≤ 2,5

Dew point temperature range

°C / °F

+5 to +94 / 41 to 201

Dew point temperature range with optional com-

pressed air dryer

°C / °F

-28 to +94 / -18.4 to 201

Max. heat compensation up to 25 °C / 77 °F at 90

% r.H.

W

400

400

1000

Max. heat compensation up to 25 °C / 77 °F at 90

% r.H. with optional compressed air dryer

W

*

*

500