6.3 Menu settings in the “User-settings” menu

User-settings

Instrument data

Contrast

Displ. Power down

Contin. operation

User Level

Safety control. Act + 100.8 °C

Safety control. Set. + 110.0 °C

29

Instrument data

Instrument Name
Enter an individual name of the alternating climate chamber.

Address
Enter a controller address (1 to 30) for operation with the communication software APT-COM™.

All other entries are relevant only for service purposes.

Contrast:
No function.

Displ. power down

Switch off event
Do not change the entry “Wait. Period”.

Waiting period
You can enter a delay time after which the display, following manual activation, will automatically be

turned off again, on condition that the moment is outside the operation time defined in menu ”Contin.

operation”.

Contin. operation
Enter an operation time to determine the period of display activity. Outside the defined period, the display

is automatically turned off. Pressing down any key will reactivate the display. After the time set in menu

“Displ. power down“, the display will turn off again as far as the actual time is not within the operation time

fixed in menu “Cont. operation“.

User Level
Toggle here to the display menu “User Level” (chap. 6.4) by entering a password. Factory default setting

for this password is +00001. You can change the password (“user code”) in the “User Level” menu.

Safety control.Set
The setting of the tolerance limit of the safety controller (chap. 12.2) is displayed. You cannot change it in

this view.

Safety control.Act
The measuring result of the safety controller is displayed. The safety controller compares the value

measured by a second independent temperature sensor to the entered tolerance limit.