MKF + MKFT (E3.1) 07/2014

page 43/114

5.5 Turning on the unit

Turn on the rear power switch (20) at least one hour before operating the unit.
Set the main power switch (3) in position I. The pilot lamp shows the unit is ready for operation.

Observe a delay time of approx. 30s between turning Off and again On the main power switch

(3). Otherwise an initialization problem might occur (display showing e.g. “–1999”).

The refrigerating and dehumidification functions are available only one hour after turning on the rear pow-

er switch (preheating phase). This is indicated by the notification “1H PREHEAT PHASE” on the controller

display. After 1 minute the message “WATER TANK EMPTY” appears on the controller display. You can

reset this message only after the 1-hour preheating phase.
Note that the chamber is in stand-by mode when the main power switch is in position I and the controller

display is dark. Turn on the unit by pressing any button. When turned on, the unit functions in the operat-

ing mode entered before turning off. In Manual Mode (HAND), the controller regulates temperature and

humidity to the last entered set-points, and in Program Mode (AUTO) it regulates temperature and humid-

ity to their set-points reached during previous program operation.

Structure of toggling between Idle Mode / Manual Mode / Program Mode:

Idle Mode

HAND

EXIT

Menu Program start

Manual Mode (HAND)

Program Mode (AUTO)

or

EXIT

For control reasons the refrigeration machine starts with a delay time. The refrigeration ma-

chine also turns off with a 5 minutes delay. This explains why the compressor may remain

operating also during positive temperature jumps.