MKF + MKFT (E3.1) 07/2014

page 54/114

9.2 Entering the temperature values and the switching states of the operation

lines

Select the first subroutine “

TP-Program 1” and confirm with ENTER.

A

program table will appear, which is initially empty until you enter the temperature values. You can now

enter the temperature program.
Temperature at the beginning

of the program section


Fan speed in %

(no function)

Program No.

Subprogram TP-Program No. 1

Total number of program

sections

Parameter set (preselected)
Tolerance band limits tem-

perature (maximum and min-

imum temperature)

Operation lines

(1= bedew protection)

Duration of program

section

Number of start section in case

of repeat cycles

Number of repeats in case of

repeat cycles

You can enter

Program sections into this program table.

Press the “PGM” button. An inquiry display appears allowing you to enter or delete individual program

sections:

ZP-Abschnitt Abs. Nr. 5

insert

delete



In this view, new program lines can be entered or deleted:

new

New lines are added below in the table

insert

New lines are added above a previously selected line

delete

Individual lines that have been selected previously are deleted

Create as many lines, i.e. program sections, as desired. As a next step, you can enter values into these

lines. It is possible to add supplementary lines later or to delete individual lines at any time.





To enter values, select the corresponding line via the

arrow keys.
Press the ENTER button. The

program editor ap-

pears.

PGM

08:43:55 15.02.02

PGM

08:43:55 15.12.13

08:43:55 15.12.13

08:43:57 15.12.13

Prog. Select.

Prog 1 PROG 01

Prog 2 PROG 02

Prog 3 PROG 03

Prog 4 PROG 04

Prog 5 PROG 05

Prog 1 PROG 01

TP- Program 2

TP- Program 3

TP- Program 1