3 entering the humidity values – BINDER MK 115 User Manual

Page 56

Advertising
background image

MKF + MKFT (E3.1) 07/2014

page 56/114

9.3 Entering the humidity values

08:43:59 15.12.13

Prog. Select.

Prog 1 PROG 01

Prog 2 PROG 02

Prog 3 PROG 03

Prog 4 PROG 04

Prog 5 PROG 05

Prog 6 PROG 06

Prog 7 PROG 07

Prog 8 PROG 08

Prog 9 PROG 09

Prog 10 PROG 10

Prog 11 PROG 11

Prog 1 PROG 01

TP- Program 2

TP- Program 3

TP- Program 1

Select the second subroutine “

TP-Program 2” and confirm by pressing ENTER. A program table will ap-

pear, which is initially empty until you enter the humidity values. You can now enter the humidity program.
Humidity at the beginning of

the program section

Fan speed in %

(no function)

Program No.

Subprogram TP-Program No. 2

Total number of program

sections

Parameter set (preselected)
Tolerance band limits humidi-

ty (maximum and minimum

humidity)

Operationlines (no function)

Duration of program

section

Number of start section in case

of repeat cycles

Number of repeats in case of

repeat cycles


Further proceeding is equivalent to the temperature value entry described in chap. 9.2.

The entry of a humidity set-point 0 % r.H. permits completely turning off humidity in defined

program sections and thus attaining faster temperature changes.


Time course of the subroutines
When starting the overall program, both subroutines (TP-Program 1 and TP-Program 2) run off synchro-

nously. They should be of the same duration because each of the subroutines becomes inactive after run-

off (i.e., no heating or refrigeration, switching states of the operation lines OFF, and the fan is off after ZP

1 is completed, no humidification after ZP 2 is completed). When the complete program is finished, the

controller changes to Idle Mode. Temperature and humidity proceed towards ambient values.

Performance after completing the program:
The controller changes to Idle Mode. Heating, refrigeration, and humidification are inactive; the chamber

approximates ambient temperature.
The fan is off. The switching states of the operation lines are OFF.

08:43:55 15.12.13