BINDER MK 115 User Manual

Page 78

Advertising
background image

MKF + MKFT (E3.1) 07/2014

page 78/114

In order to ensure accurate humidifying, observe the following points with regard to the

freshwater supply:

Supply pressure 1 to 10 bar when connecting to a water pipe

Water type: deionized (demineralized) water.

To ensure humidification during 24 hours even at high humidity set-points with manual wa-

ter supply, we recommend filling the freshwater can at the end of each day.

Water intake temperature NOT below +5 °C / 41°F and not exceeding 40 °C / 104°F.

BINDER GmbH is NOT responsible for the water provided by the customer.
Any problems and malfunctions that might arise following use of water of deviating quality is

excluded from liability by BINDER GmbH.

Automatic fresh water supply via water pipe
With this type of supply, the humidity system is continuously functional. The correct water supply can be

monitored at the internal water can, where the water is intermediately stored also if a water pipe is con-

nected. The correct filling level is automatically maintained at ½ to ¾ of the maximum level.
Manual fresh water supply via internal freshwater can
With this type of supply, the humidity system is functional only if the water can is sufficiently filled. Check

the filling level daily. The water reserve in the water can is sufficient for for a period, which may last be-

tween one and several days, depending on the humidity demand (entered humidity set-point and number

of door openings). Fill the can up to the maximum level mark only. The cover of the water inlet valve must

be screwed on the freshwater connection “IN” (18) (chap. 4.2.2).
Wastewater
The condensation water from the interior and excess freshwater (by manual excess filling or in case of

fault) is collected in an internal can with a volume of approx. 1.5 liters. It is pumped off via the wastewater

pipe.
Dehumidifying system
When switch Humidity ON / OFF (4) (located on the lateral control panel) is in position ON, the alternating

climate chamber MKF / MKFT dehumidifies as needed in order to reach the entered humidity set-point

inside the Control range of temperature and relative humidity (Figure 19).
Dehumidification occurs in case of need by means of a defined dew point undershoot of several evapora-

tors of the refrigeration system. The condensate which forms is carried away as wastewater.
With temperature set-points outside the control range (hatched area in Figure 19), humidification and

dehumidification are automatically turned off. If the humidity system is turned off while there are descend-

ing temperature curves, then operation of the refrigeration system may cause dehumidification of the

charging material.
With humidity set-points outside the control range (hatched area in Figure 19), or with entry of set-point 0

% r.H., the humidification and dehumidification system is turned off even if the humidity switch is in posi-

tion ON.
For error indications concerning water supply and humidity system, see chap.13.1 and 19