Notification and alarm functions – BINDER MK 115 User Manual

Page 74

Advertising
background image

MKF + MKFT (E3.1) 07/2014

page 74/114

13. Notification and alarm functions

13.1 Notification and alarm system overview (auto diagnosis system)

Visual indications of notification or error messages are blue notes on the display of the MB1controller.

Visual indications of an alarm messages are red notes with an alarm bell symbol.

In addition, there is an audible alert, if you did not deactivate the buzzer in the “User level” menu

(chap. 6.4).

Event

Note

(blue field)

Alarm

(red field)

Lack of water in the freshwater can. The unit

continues working for a few hours.
In case of freshwater supply via water pipe,

the water tap is closed, or the unit is defective.

WATER LOW

after 5 min

Wastewater pump or float switch in the

wastewater can defective (contact BINDER

service), or wastewater hose clogged.

WASTEWATER

after 10 sec.

Humidifying unit defective. Contact BINDER

service.

FAULT STEAM SYSTEM

immediately

Freshwater can is too empty to allow normal

function, or the unit is defective.
The humidification system turns off.
Fill up the water can or open the water supply.
During the one-hour preheating phase: mes-

sage without significance

WATER TANK EMPTY

after 60 sec.

When manually filling up the freshwater can:
Freshwater can is too full, or the unit is defec-

tive. Suck off the water (chap. 4.2.2). The unit

functions as usual.
If the message persists, contact BINDER ser-

vice.

WATER LEVEL TOO

HIGH

after 60 sec.

Humidity switch OFF or humidity outside con-

trol range. No active humidification of the inte-

rior. Further dehumidification possible by re-

frigerating operation

HUMID OFF

immediately

One-hour preheating phase, no refrigerating

and dehumidification functions

1H PREHEAT PHASE

immediately

Fault in refrigerating machine. Contact

BINDER service.

FAULT COMPRESSOR

immediately