page 57/114

9.4 Selecting between set-point ramp and set-point step

Set-points always refer to the start of a program section, i.e., at the beginning of each program section

the entered set-point is targeted. During program section operation, the temperature or humidity gradually

passes to the set-point entered for the next program section.
By appropriate planning of the program section timing, you can enter all kinds of temperature and humidi-

ty transitions.

Gradual temperature / humidity changes “set-point ramp”
The set-point changes its value gradually while proceeding from one program section to the next one

during the programmed section length. The actual temperature or humidity value (X) follows the con-

tinually moving set-point (W) at any time.

Program sections with constant temperature / humidity
The initial values of two subsequent program sections are identical; so the temperature or humidity

remains constant during the whole time of the first program section.

Sudden temperature / humidity changes “set-point step”
Steps are temperature or humidity changes (ramps) that occur during a very short interval. A section

with a different set-point follows two program sections with an identical set-point. If the duration of this

transitional program section is very short (minimum entry 1 sec), the temperature or humidity change

will proceed rapidly within the minimum amount of time.

Figure 17: Possible temperature or humidity transitions


The following chapter offers examples of programming a set-point ramp and a set-point step.

9.5 Program entry as set-point ramp or as set-point step

In order to avoid incorrect programming, we recommend plotting both the temperature and humidity pro-

files (chart templates in chap. 9.11 and 9.12) and entering the values into a table (templates in chap. 9.13

and 9.14).
The controller provides 8 operation lines that can be activated or de-activated for each program section.

Operating contact 1 can be used to control the bedew protection (chap. 10). Operation lines 2 to 5 serve

to switch any device connected to the zero-voltage relay outputs (DIN sockets (11) and (12), chap. 11).

Operation line 8 releases the compressed air dryer (option, chap. 16.7). The other operation lines are

non-functional.

01

02

03

04

05

06

07

08

09

W

t