MKF + MKFT (E3.1) 07/2014

page 80/114

16. Options

16.1 Communication software APT-COM™ 3 DataControlSystem (option)

The alternating climate chamber is regularly equipped with an Ethernet interface (10a) that can connect

the BINDER communication software APT-COM™ 3 DataControlSystem can be connected. The MAC

Address is indicated next to the Ethernet interface. The actual temperature and humidity values are given

at adjustable intervals. Programming can be performed graphically via PC. Up to 30 chambers with RS

422 interface can be cross-linked. For further information, please refer to the operating manual of the

BINDER communication software APT-COM™.

16.2 Interface RS 422 (option)

With this option, the chamber is equipped with a serial interface RS 422 (10b) instead of the Ethernet

interface (10a) that can connect the BINDER communication software APT-COM™ 3 DataControlSys-

tem. The actual temperature and humidity values are given at adjustable intervals. For further information,

please refer to the operating manual of the BINDER communication software APT-COM™.
Pin allocation of the RS 422 interface: Pin 2:

RxD (+)

Pin 3:

TxD (+)

Pin 4:

RxD (-)

Pin 5:

TxD (-)

Pin 7:

Ground

16.3 Analog outputs for temperature and humidity (option)

With this option, the chamber is equipped with analogue outputs 4-20 mA for actual value and set-point

value of temperature and of humidity. These outputs allow transmitting data to external data registration

systems or devices.
The connection “analog outputs for temperature” is realized as a DIN socket (13) in the lateral control

panel as follows:

ANALOG OUTPUT TEMPERATURE 4-20 mA DC
PIN 1: Temperature actual value –

PIN 2: Temperature actual value +

PIN 4: Temperature set-point value –

PIN 5: Temperature set-point value +

MKF: Temperature range: -40 °C / -40°F up to +180 °C / -356°F

MKFT: Temperature range: -70 °C / -94°F up to +180 °C / -356°F

A suitable DIN plug is enclosed.

Figure 20: Pin allocation of DIN socket (13) for option analog outputs for temperature
The connection “analog outputs for humidity” is carried out as a DIN socket (14) in the lateral control pan-

el as following:

ANALOG OUTPUT HUMIDITY 4-20 mA DC
PIN 1: Humidity actual value –

PIN 2: Humidity actual value +

PIN 4: Humidity set-point value –

PIN 5: Humidity set-point value +

Humidity range: 0 % r.H. up to 100 % r.H.

A suitable DIN plug is enclosed.