MKF + MKFT (E3.1) 07/2014

page 61/114


Program table corresponding to the diagram above:

Program

section

No.

Set-

point

temp.

W-1

Fan

FAN

Section

time

Time

Operation lines

Sk

Target

section

No

No. of

cycles

Cy

Minimum

tolerance

Tmin

Maximum

tolerance

Tmax

8 7 6 5 4 3 2 1

01

-20

****. 00:30:00 0 0 0 0 0 0 0 0

1

0

-1999

+9999

02

-20

****. 00:00:01 0 0 0 0 0 0 0 0

1

0

-1999

+9999

03

0

****. 01:30:00 0 0 0 0 0 0 0 0

1

0

-5

+5

04

0

****. 00:00:01 0 0 0 0 0 0 0 0

1

0

-1999

+9999

05

60

****. 01:00:00 0 0 0 0 0 0 0 0

1

0

-2

+2

06

60

****. 00:00:01 0 0 0 0 0 0 0 0

1

0

-1999

+9999

07

-40

****. 03:20:00 0 0 0 0 0 0 0 0

1

0

-1999

+9999

08

-40

****. 00:00:01 0 0 0 0 0 0 0 0

1

0

-1999

+9999

Now enter the values of the above program table into one of the 25 program places of the controller MB1:

For rapid transition phases, do NOT program any tolerance limits in order to permit maximum

heating, refrigerating, and humidification speed.