BINDER MK 115 User Manual

Page 55

Advertising
background image

MKF + MKFT (E3.1) 07/2014

page 55/114


Enter the individual values of the selected program section.

Program editor Abs.Nr. 6

****.*

Setpoint 1

+100.0

FAN

Operating contacts

Time

00:45:00

Repeat Section

5

Repeat Number

10

Tol.-band min.

-1999.0

Tol.-band max.

+9999.0

Parameter set

1

Setpoint 1

Temperature at the start of the program section

FAN

Fan speed in % (no function)

Operating contacts Operation lines ON/OFF

(chap. 10, 11, and 16.7)

Time

Duration of the program section

Repeat Section

No. of start section in case of repeat cycles

Repeat Number

No. of repeats in case of repeat cycles

Tol.-band min.

Temperature limits (maximum / minimum temperature)

(In case of exceeding: temporary program stop)

Tol.-band max.
Parameter set

Pre-selected value (Do NOT change!)

Select the parameters via the arrow keys and confirm by pressing ENTER.
Then enter the values via the arrow keys, and confirm the entry by pressing ENTER.

For a negative set-point entry, enter the numerical value first, and then the minus sign (-).

With set-point type “

Limit”, the user shall adapt the safety controller (chap. 12.2) to the high-

est temperature set-point value of the program actually used. Check the safety controller for

each temperature program and change it if necessary. Set the safety controller set-point by

approx 10 °C above the highest temperature set-point of the program.

In case of the optional over-/under temperature safety device (chap. 12.3), check also the

temperature limits entered there, and adjust them if necessary.

Performance after completing the program:
The controller changes to Idle Mode. Heating, refrigeration, and humidification are inactive; the chamber

approximates ambient temperature.
The fan is off. The switching states of the operation lines are OFF.