Configuring the HeartStart XLT
10-10
Setting Up and Configuring the HeartStart XLT
3ULQWHU 'HOD\
Captures what you just saw. All printed
strips, including those generated by an
event (mark, charge, shock or alarm),
include an additional 6 seconds of infor-
mation - the 6 seconds of information that
occurred just prior to printing being initi-
ated.
6 Sec Delay /
No Delay
3DFH 3XOVH 0DUNHUV
Shows pace pulse markers on the ECG
displayed, if an internal pacemaker is
detected.
Leads & Pads /
Leads Only
(&* /HDG &DEOH
Selects the monitoring electrodes source.
3-lead / 5-lead
3RZHU RQ 0RGH
Defines the operating mode when the
device is turned on. Applies to a new use,
not to a continued use.
AED / Manual
456 %HHSHU
Provides an audible beep with each QRS
complex detected.
Manual / AED
& Manual / Off
Table 10-1 General Settings (Continued)
Parameter
Description
Setting
Choices