Operational Checks

11-2

Maintaining the HeartStart XLT

Operational Checks

The following operational checks are intended to quickly verify the proper
operation of the HeartStart XLT. Perform these checks regularly, at the inter-
vals specified, along with visual inspection of the device and all cables, con-
trols, accessories and supplies. Also regularly check expiration dates of all
supplies, such as multifunction defib electrode pads and monitoring
electrodes.

Before You Begin

Before you run the Shift/System Check, be aware of the following conditions:

z

Do not touch any of the controls on the HeartStart XLT while the
Shift/System Check is running.

z

If a

)DLOXUH or 6HUYLFH 8QLW message is displayed, or if an unexpected 1RW

7HVWHGresult is displayed, check that the test is set up correctly. Make
sure that:
‰ the paper is in the printer
‰ the test load is attached
‰ a Data Card with enough space is inserted into the HeartStart

XLT

‰ a charged battery is inserted into the HeartStart XLT
Run the Shift/System Check again, ensuring that no one touches any
of the controls on the defibrillator unless prompted to do so.

z

If the Data Card is full, the message

6HUYLFH 8QLW appears at the bottom

of the screen and the message

'DWD &DUG )XOO appears at the top of the

screen. Replace the Data Card and perform the Shift/System Check
again. If the message

6HUYLFH 8QLW continues to appear, do not use the

device and call for service.