Specifications

13-6

Specifications & Safety

Thermal Array Printer (cont.)

Auto Printing: The printer can be configured to automatically print on Mark,
Charge, Shock and Alarm.

Delayed Printing: The printer can be configured to run real time or with a six
second delay.

Reports: The following can be printed: Event Summary, Configuration,
Extended Self Test, System Log.

Speed: 25 mm/s with an accuracy of

5%.

Amplitude Accuracy:

10% or

50 uV, whichever is greater.

Paper Size: 50 mm by 30 m (100 ft).

Battery and Battery Power Module

Battery Type: 2 Ah, 12V, rechargeable, Sealed Lead Acid.

Battery Dimensions: 2.4" (H) x 0.94" (W) x 7.2" (D).

61.7 mm (H) x 23.9 mm(W) x 182 mm (D).

Weight: 1.4 lb (0.65 kg).

Charge Time:

Approximately 14.5 hours to 100%, indicated by LED on power module.
Approximately 2.5 hours to 90%, indicated by LED on power module.

Capacity: 2.7 hours ECG monitoring or 50 full-energy discharges or 2.1
hours ECG monitoring while pacing (with a new, fully charged battery and

temperature above 20

o

C).

Battery Indicators: LOW BATTERY message appears on monitor when at
least 10 minutes of monitoring time and 6 maximum-energy discharges

remain (with a new battery and temperature above 20

o

C).

Battery Storage: Should not be stored above 40

o

C for extended periods of

time.

Charger Output: Unit can be operated using Class I AC power module alone
with no battery installed.

+