Troubleshooting, Troubleshooting -10 – Philips HEARTSTART XLT M3500B User Manual

Page 70

Advertising
background image

Troubleshooting

5-10

Monitoring SpO

2

Troubleshooting

The table below lists system messages that you may encounter when monitor-
ing SpO

2.

Table 5-2 System Messages

Problem or Message

Possible Cause

Corrective Action

6S2 1RQ 3XOVDWLOH

z

Pulse absent or too
weak to be detected.

z

Check the sensor is
applied properly.

z

Make sure the sensor
site has a pulse.

z

Relocate the sensor to
a site with improved
circulation.

z

Try another sensor
type.

6S2 /RZ 6LJQDO

z

SpO

2

signal is too low

to give an accurate read-
ing.

z

Check the sensor is
applied properly.

z

Try another sensor
type.

6S2 1RLV\ 6LJQDO

z

Excessive patient move-
ment, electrical interfer-
ence, or optical
interference.

z

Minimize patient
motion or apply sen-
sor to site with less
movement.

z

Secure the sensor
cable loosely to the
patient.

z

Reduce sources of
electrical or optical
interference.

6S2/LJKW ,QWHUI

z

The level of ambient
light is so high that the
sensor cannot obtain an
SpO

2

reading.

z

Sensor or cable is dam-
aged.

z

Cover sensor with an
opaque material.

z

Check sensor for dam-
age; try another sen-
sor.