Defibrillating

M3500B HeartStart XLT Defibrillator/Monitor

6-5

6

Figure 6-4 Charging Display

If desired, you may increase or decrease the selected energy level after

pressing the

button. To do so, press

repeatedly until the

desired energy level is displayed. The defibrillator charges to the selected
energy automatically. Wait until the current charge reaches the selected
energy level before proceeding.

3. Shock

Confirm that a shock is still indicated. Make sure no one is touching the
patient or anything connected to the patient. Call out “Clear.” Then press

to deliver a shock to the patient.

To disarm the defibrillator, press

. If

is not pressed in 30 sec-

onds, the defibrillator disarms automatically.

If additional shocks are indicated, repeat the defibrillation process

WARNING

Defibrillation current can cause operator or bystander injury. Do not touch the
patient, or equipment connected to the patient, during defibrillation.

ENERGY

SELECT

HR ALARM

LEAD

SELECT

DISARM

Shocks:0

00:00:8

78

Pads

SpO2

150 J

87 J

Current Charge

&+$5*(

(1(5*<

6(/(&7

6+2&.

',6$50

6+2&.