Operational Checks

11-6

Maintaining the HeartStart XLT

Power Supply

Battery - make sure:

z

a charged battery is in the HeartStart XLT

z

another battery is charged or being charged

z

the batteries have no visible signs of damage

AC/Power Module

1. Make sure a battery is in the HeartStart XLT.

2. Plug the power module into a power outlet and connect it to the Heart-

Start XLT.

3. Verify that the power and charging indicators on the power module are

lit.

4. Remove the battery from the HeartStart XLT and verify that the charg-

ing indicator on the power module is no longer lit. Replace the battery.

Printer - make sure the printer:

z

has sufficient paper

z

prints properly