Delivering a Synchronized Shock

M3500B HeartStart XLT Defibrillator/Monitor

7-3

7

Figure 7-1: Charging in Sync Mode

If desired, you may increase or decrease the selected energy level after

pressing

,

by pressing

repeatedly until the desired energy

level is displayed. The defibrillator charges to the modified energy level
automatically. Wait until the current charge reaches the selected energy
level before proceeding.

5. Make sure no one is touching the patient or anything connected to the

patient. Call out “Clear.” Then press

and continue to hold

down until the shock is delivered.

The defibrillator shocks with the next detected R-wave.

WARNING

Defibrillation current can cause operator or bystander injury. Do not touch the
patient, or equipment connected to the patient, during defibrillation.

ENERGY

SELECT

HR ALARM

LEAD

SELECT

DISARM

Shocks:0

00:00:8

78

Pads

SpO2

150 J

87 J

Sync

&+$5*(

(1(5*<

6(/(&7

6+2&.

6+2&.