Philips HEARTSTART XLT M3500B User Manual

Page 142

Advertising
background image

System Messages

12-6

Troubleshooting

'DWD &DUG )XOO

z

The incident is more than 2 hours in
duration, causing the Data Card to fill.

z

An empty Data Card was not inserted
for the incident, causing the Data Card
to fill sooner.

z

None. A new Data Card can not be
inserted during an incident.

z

Use one Data Card per incident/
patient to decrease the chance of the
card filling.

'DWD &DUG ,QWHUUXSWHG

The Data Card is removed during an inci-
dent.

z

None. The Data Card can not be re-
inserted during an incident.

z

Do not remove the Data Card during
an incident.

'DWD &DUG 1RW ,Q 6HUYLFH

The Data Card is inserted while the Heart-
Start XLT is on.

None. A Data Card must be inserted prior
to turning the HeartStart XLT on for the
current patient.

,QFRPSDWLEOH 'DWD &DUG

A Data Card other than the M3510A is
inserted.

Use only M3510A Data Cards.

1R 'DWD &DUG 3UHVHQW

A Data Card is not in the HeartStart XLT.

Turn the HeartStart XLT off and insert a
Data Card prior to the first event for the
patient.

.H\ ,QDFWLYH

The key pressed is currently inactive (i.e.

and

are inactive in

AED Mode).

Use the appropriate mode for the key.

6WRS 3DFHU

is pressed while pacing pulses

are being delivered.

Stop pacing before changing the pacing
mode.

Table 12-2 Momentary Messages (Continued)

Message

Possible Cause

Corrective Action

Sync

Pacer

Mode