Troubleshooting Tips

12-8

Troubleshooting

There is a dashed (----)
line on the display
instead of an ECG.

ECG data is not being acquired.

z

Check the patient cable is connected.

z

Check the pads or electrodes are prop-
erly applied.

z

Check that the desired lead is selected.

The HeartStart XLT
does not appear to be
functioning properly.

z

The battery is low.

z

There is a system failure.

z

Change the battery.

z

Take the device out of use and call for
service.

The displayed time is
incorrect.

The time was not correctly set in the con-
figuration.

Set the time in the General Settings menu
of the Configuration Mode.

The printed date is
incorrect.

The date was not correctly set in the con-
figuration.

Set the time in the General Settings menu
of the Configuration Mode.

Table 12-3 Troubleshooting Tips (Continued)

Situation

Cause

Possible Solution