Setting the heart rate alarm, Setting the heart rate alarm -6 – Philips HEARTSTART XLT M3500B User Manual

Page 58

Advertising
background image

Setting the Heart Rate Alarm

4-6

Monitoring the ECG

Setting the Heart Rate Alarm

The computed heart rate (number of detected QRS complexes per minute) is

displayed below the

sofkey, next to the

. The heart rate represents

the number of QRS complexes detected in a minute. A QRS beeper, if config-
ured to on, indicates each QRS complex detected.

If desired, a HR alarm may be set to alert you when the heart rate is outside
the specified limits. Limit choices are listed in Table 4-4.

To set a HR alarm, cycle through the limit choices by pressing

until

the desired limits are shown. The

then appears next to the heart rate value

to indicate that the HR alarm is set.

WARNING

Heart rate alarms are temporarily suspended in AED Mode during ECG analysis or

when

is pressed (for the duration of the paused period). Heart rate alarms are

also suspended while charging for defibrillation and delivering a shock.

WARNING

Heart rate displays and alarms function with internal and external pacemakers, but
they can be unreliable. Observe the patient closely if pacemakers are used.

Table 4-4 HR Alarm Limit Choices

Alarm If Over:

Or under:

100

30

140

60

160

90

200

120

+5 $/$50

+5 $/$50