ECG Monitoring

Inputs: Single channel ECG may be viewed on display and printed. PADS
ECG is obtained through two multifunction defibrillation electrode pads.
Lead I, II, or III is obtained through the 3-lead ECG cable and separate moni-
toring electrodes. With a 5-lead cable, lead aVR, aVL, aVF, or V can also be
obtained.

Lead Fault: LEADS OFF message and dashed line appear on the monitor if a
lead electrode or wire becomes disconnected.

Pad Fault: PADS OFF message and dashed line appear on the monitor if a
pad becomes disconnected.

Heart Rate Display: Digital readout on display from 15 to 300 bpm, with an

accuracy of

10%.

Heart Rate Alarms: Configurable pairs of low and high heart rate alarm lim-
its: 30 to 100, 60 to 140, 90 to 160, and 120 to 200 bpm.

Defibrillation Patient Cable Length: 7 ft (2.13 m).

ECG Cable Length: 10 ft (3.05 m).

Common Mode Rejection: Greater than 90 dB measured per AAMI stan-
dards for cardiac monitors (EC 13).

ECG Size: 2.5, 5, 10, 20, 40 mm/mV.

Frequency Response:

AC Line Filter: 60

Hz/50Hz.

Pads ECG for Display:

Monitor (.15-40 Hz)/EMS (1-30 Hz).

Pads ECG for Printer: Monitor

(.15-40

Hz)/EMS (1-30 Hz).

Leads ECG for Display: Monitor

(.15-40

Hz)/EMS (1-30 Hz).

Leads ECG for Printer:

Diag (.05-150 Hz)/EMS (1-30 Hz)/
Monitor (.15-40 Hz).

Patient Isolation (defibrillation proof):

ECG: Type CF
SpO

2

: Type CF

Defib: Type BF (multi-function defib electrode pads)

+