Preparing for Pacing

M3500B HeartStart XLT Defibrillator/Monitor

8-3

8

Preparing for Pacing

In preparation for pacing:

1. Apply multifunction defib electrode pads as directed on the package. Use

either the anterior-anterior or anterior-posterior placement, as appropriate.

2. Connect the pads to the patient cable. (See Figure 6-2.)

3. If needed, insert a Data Card (as described in “Using a Data Card” on 2-

11).

4. Press On.

5. Enable Manual Mode. (See “Enabling Manual Mode” on page 6-2.)

In addition, for demand mode pacing:

1. Apply monitoring electrodes. (See “Applying Monitoring Electrodes” on

page 4-2.)

2. Use

to select the best lead with an easily detectable R-wave (See

“Selecting the Lead” on page 4-5.) If you do not select a lead (i.e. pads is
the selected ECG source),

/HDG , is automatically selected when the pacing

function is turned on.

NOTE

If pacing for long periods of time, new monitoring electrodes and multifunc-
tion defib electrode pads may need to be applied periodically. Refer to the
manufacturer’s documentation for how often to change monitoring electrodes
or defib pads.

/($'